MP150 Series

Product Details

  • Safety Approvals: Medical 60601-1
  • Footprint: 5 x 3 x 1.4”
  • Construction: Class I or Class II
  • High Efficiency: 90% Typical At Full Load
  • Output: 5VSDB @ 0.5A
  • Input: Universal AC Input With Active PFC
  • Emissions: EN55011 / FCC Part 15J Class B
  • Warranty: 3 Years
  • Additional Feature: Remote On/Off, Power Fail, Remote Sense

150 Watt Medical Desktop Power Supply

Reliability and value, the 150W 3 x 5″ Open Frame is the industry-standard workhorse. Phasium’s high-efficiency PF + LLC put’s out less heat and has inherently low EMI. Features include a 12V fan output and remote on/of. Class I and Class II construction is available.

MODELS MAIN OUTPUT
VOLTAGE
OUTPUT CURRENT (Convection) OUTPUT CURRENT (FORCED AIR) STANDBY OUTPUT FAN OUTPUT
MP150-12 12V  12.5A 16.7A 5V 400mA 12V 0.4A
MP150-19 19V  7.9A 10.5A 5V 400mA 12V 0.4A
MP150-24 24V  6.3A 8.3A 5V 400mA 12V 0.4A
MP150-28 28V  5.4A 7.1A 5V 400mA 12V 0.4A
MP150-48 48V  3.1A 4.2A 5V 400mA 12V 0.4A

Input Voltage– 90-264 VAC (Universal)

Efciency- 90% typical at 115/230 VAC, full load

Leakage Current– <100µA @ 264VAC

Operating Temperature– 0°C ~ +40°C

Protection– Internal T5A / 250V Fuse (Line & Neutral)

Line Regulation– ±0.5%

Isolation- Input to Output: 4000 VAC

Input to Ground: 1500 VAC

Output to Ground: 500 VDC

Load Regulation– ±2% For Main & 5VSDB ; ±10% For Fan

MTBF– 300K Hours, Mil-217F at 25 °C, full load

Download Product Summary Sheet

Contact us to find out more about this revolutionary product